12 Channels Parallel Cuvette Cells | 410 nm LED Solid-State Optics | < 0.5% Repeatability Precision | 10.1" Touch Android 11 Smart OS |
The FSTest PR12 Rapid Pesticide Residue Detector utilizes spectrophotometric determination based on the selective absorption of visible light by target chemical molecules, strictly following the Lambert-Beer Law (A = -lg(I/I₀) = -lgT = klc). By comparing the absorbance of unknown agricultural samples against known reference standards, the instrument performs rapid qualitative screening and quantitative inhibition rate calculation.
Engineered with an all-solid-state 12-channel optical architecture, fixed-position absorption cells, a 10.1-inch HD capacitive touchscreen running Android 11, and an integrated thermal printer, the PR12 delivers high-throughput, field-ready pesticide residue screening for agricultural production bases, wholesale produce markets, supermarket QA departments, and food safety regulatory authorities.
High-Throughput 12-Channel Parallel Testing: Equipped with 12 fixed-position colorimetric channels (1 cm optical path), enabling simultaneous batch screening of up to 12 samples with automatic channel sensing.
All-Solid-State High-Precision Optics: Utilizes a built-in 410 nm discrete-wavelength high-brightness LED light source with zero mechanical moving parts, achieving transmittance repeatability < 0.5% and photocurrent drift ≤ ±1%.
Intelligent Auto-Subtraction & Instant Calculation: Features one-touch "Auto-subtract" mode that automatically assigns the control blank and calculates enzyme inhibition rates against built-in national food safety standards (e.g., GB 5009.199).
Standalone Smart Android 11 Interface: Operated via a 10.1-inch (1280×800) touchscreen (with mouse and keyboard support), complete with onboard digital manuals, instructional videos, and 32 GB local data storage.
Built-In Thermal Printer & Full Connectivity: Prints instant benchtop test receipts automatically upon completion and supports WiFi, Bluetooth, RJ45 Ethernet, HDMI, and USB data export.
| Product Name | PR12 Rapid Pesticide Residue Detector |
|---|---|
| Analytical Principle | Visible Spectrophotometry / Enzyme Inhibition Rate Method (Lambert-Beer Law) |
| Optical System & Light Source | All-solid-state multi-channel optical path; Built-in 410 nm discrete-wavelength high-brightness LED |
| Colorimetric Cell Configuration | 12 parallel channels, fixed-position absorption cell, 1 cm optical path |
| Optical Accuracy Metrics | Transmittance repeatability: < 0.5% | Photocurrent drift: ≤ ±1% | Transmittance error: ≤ ±2% |
| Processor & Display | Quad-core ARM CPU | 2 GB RAM | 32 GB Storage (MicroSD expandable) | 10.1-inch Touchscreen (1280×800) |
| Operating System & Input | Android 11 OS; Supports Touch, USB Mouse, and Keyboard operation |
| Ports & Wireless Connectivity | 2× USB 3.0 Type-A, 2× USB 2.0 Type-A, 1× USB 2.0 Type-B, HDMI, RJ45 Ethernet, MicroSD slot, Wi-Fi, Bluetooth |
| Printer & Power Supply | Built-in thermal printer; Power supply: 12 V DC, 5 A (Adapter input: 100–240 V AC, 50/60 Hz) |
| Optical Parameter | Technical Specification | Performance Advantage |
|---|---|---|
| Peak Wavelength | 410 nm Discrete LED | Optimized for cholinesterase inhibition assays |
| Channel Capacity & Path | 12 Channels / 1 cm Optical Path | Simultaneous multi-sample batch screening |
| Transmittance Repeatability | < 0.5% | High measurement consistency across runs |
| Photocurrent Drift & Error | Drift ≤ ±1% | Error ≤ ±2% | Long-term baseline stability without warm-up lag |
Ideal for rapid screening of organophosphate and carbamate pesticide residues across agricultural and food supply chains:
STEP 1 Login & Setup Power on and log into the PR12 App. Tap "Quick Test" → "Spectrophotometry", then select the detection method, sample names, and testing organization. | STEP 2 Load Cuvettes Tap "Clear" to reset channel data. Open the chamber lid and insert colorimetric cuvettes (Control in Ch. 1, Samples in Ch. 2–12). The system auto-detects active cells. | STEP 3 Auto-Subtract Read Tap "Auto-subtract" (or "Sample") to initiate reading. The system deducts the blank control automatically and computes inhibition rates after reaction countdown. | STEP 4 Print & Archive Review pass/fail compliance judgments on screen. Tap "Print" for instant thermal receipt output or enter "Detection Record" to query and export historical logs. |
Auto-Subtract Mode: Automatically designates the first inserted channel as the Control (C) blank and remaining active channels as Samples (S), streamlining batch enzyme inhibition testing.
Individual Channel Control: Operators can also click individual [C] or [S] icons beside any of the 12 channels to run independent control baselines or single-channel readings.
Automated Standard Judgment: Built-in regulatory limits (e.g., inhibition rate < 50% threshold per GB 5009.199) automatically flag samples as compliant (OK) or exceeding limits.
Adhere to standard instrument maintenance guidelines to guarantee long-term optical precision:
Cuvette Removal Before Moving: Always remove all colorimetric cuvettes from the detection chamber immediately after testing to prevent liquid spills from damaging internal optics during transport.
Operating Environment: Install on a level, vibration-free benchtop (away from centrifuges or strong magnetic fields) at 5°C–35°C and relative humidity < 70%, out of direct sunlight.
Optical Surface Care: Keep cuvette optical faces clean and free of fingerprints or scratches. Never spray liquid cleaners directly into the detection chamber or ports.
Dedicated Power Supply: Use exclusively the supplied 12 V DC / 5 A power adapter connected to a grounded 100–240 V AC outlet.
Instrument Package Included: 1× PR12 Rapid Pesticide Residue Detector Main Unit, 1× Power Adapter (12 V DC, 5 A) & Power Cord, Colorimetric Cuvettes, Thermal Printing Paper Roll, User Manual.
Compatible Consumables & Accessories (Ordered Separately): FSTest Pesticide Residue Rapid Test Kits (Enzyme Inhibition Reagents), replacement 1 cm colorimetric cuvettes, thermal printer rolls, and USB mouse/keyboard.