home_icon
Home
>
Products
>
Pesticide Residue Detector
>
PR12 Rapid Pesticide Detector
PR12 Rapid Pesticide Detector
Brand: FSTest
CatNO: 50012
Application: Rapid detection of organophosphorus and carbamate pesticide residues in fruits, vegetables, tea and produce
PR12 Rapid Pesticide Detector
PR12 Rapid Pesticide Detector

PR12 Rapid Pesticide Residue Detector

12-Channel Spectrophotometric Analyzer for Food & Agricultural Safety
ANALYTICAL INSTRUMENTATION
       Model: PR12 | Brand: FSTest
12 Channels
Parallel Cuvette Cells
410 nm LED
Solid-State Optics
< 0.5%
Repeatability Precision
10.1" Touch
Android 11 Smart OS
1. Product Overview

The FSTest PR12 Rapid Pesticide Residue Detector utilizes spectrophotometric determination based on the selective absorption of visible light by target chemical molecules, strictly following the Lambert-Beer Law (A = -lg(I/I₀) = -lgT = klc). By comparing the absorbance of unknown agricultural samples against known reference standards, the instrument performs rapid qualitative screening and quantitative inhibition rate calculation.

Engineered with an all-solid-state 12-channel optical architecture, fixed-position absorption cells, a 10.1-inch HD capacitive touchscreen running Android 11, and an integrated thermal printer, the PR12 delivers high-throughput, field-ready pesticide residue screening for agricultural production bases, wholesale produce markets, supermarket QA departments, and food safety regulatory authorities.

2. Key Benefits
  • High-Throughput 12-Channel Parallel Testing: Equipped with 12 fixed-position colorimetric channels (1 cm optical path), enabling simultaneous batch screening of up to 12 samples with automatic channel sensing.

  • All-Solid-State High-Precision Optics: Utilizes a built-in 410 nm discrete-wavelength high-brightness LED light source with zero mechanical moving parts, achieving transmittance repeatability < 0.5% and photocurrent drift ≤ ±1%.

  • Intelligent Auto-Subtraction & Instant Calculation: Features one-touch "Auto-subtract" mode that automatically assigns the control blank and calculates enzyme inhibition rates against built-in national food safety standards (e.g., GB 5009.199).

  • Standalone Smart Android 11 Interface: Operated via a 10.1-inch (1280×800) touchscreen (with mouse and keyboard support), complete with onboard digital manuals, instructional videos, and 32 GB local data storage.

  • Built-In Thermal Printer & Full Connectivity: Prints instant benchtop test receipts automatically upon completion and supports WiFi, Bluetooth, RJ45 Ethernet, HDMI, and USB data export.

3. Technical Specifications & Optical Performance
Product NamePR12 Rapid Pesticide Residue Detector
Analytical PrincipleVisible Spectrophotometry / Enzyme Inhibition Rate Method (Lambert-Beer Law)
Optical System & Light SourceAll-solid-state multi-channel optical path; Built-in 410 nm discrete-wavelength high-brightness LED
Colorimetric Cell Configuration12 parallel channels, fixed-position absorption cell, 1 cm optical path
Optical Accuracy MetricsTransmittance repeatability: < 0.5% | Photocurrent drift: ≤ ±1% | Transmittance error: ≤ ±2%
Processor & DisplayQuad-core ARM CPU | 2 GB RAM | 32 GB Storage (MicroSD expandable) | 10.1-inch Touchscreen (1280×800)
Operating System & InputAndroid 11 OS; Supports Touch, USB Mouse, and Keyboard operation
Ports & Wireless Connectivity2× USB 3.0 Type-A, 2× USB 2.0 Type-A, 1× USB 2.0 Type-B, HDMI, RJ45 Ethernet, MicroSD slot, Wi-Fi, Bluetooth
Printer & Power SupplyBuilt-in thermal printer; Power supply: 12 V DC, 5 A (Adapter input: 100–240 V AC, 50/60 Hz)
Spectrophotometric Optical Module Specifications Summary
Optical ParameterTechnical SpecificationPerformance Advantage
Peak Wavelength410 nm Discrete LEDOptimized for cholinesterase inhibition assays
Channel Capacity & Path12 Channels / 1 cm Optical PathSimultaneous multi-sample batch screening
Transmittance Repeatability< 0.5%High measurement consistency across runs
Photocurrent Drift & ErrorDrift ≤ ±1% | Error ≤ ±2%Long-term baseline stability without warm-up lag
4. Applicable Sample Types & Industry Sectors

Ideal for rapid screening of organophosphate and carbamate pesticide residues across agricultural and food supply chains:

Fresh Vegetables & Leafy Greens (Tomato, potato, lettuce, cabbage, brassicas)    Orchard Fruits & Melons (Apple, citrus, berries, stone fruits)    Tea Leaves, Grains & Agricultural Raw Materials    Wholesale Produce Markets & Supermarket Distribution Centers    Agricultural Farms, Catering Central Kitchens & Regulatory Inspection Bureaus
5. Instrument Operation Procedure
STEP 1
Login & Setup
Power on and log into the PR12 App. Tap "Quick Test" → "Spectrophotometry", then select the detection method, sample names, and testing organization.
STEP 2
Load Cuvettes
Tap "Clear" to reset channel data. Open the chamber lid and insert colorimetric cuvettes (Control in Ch. 1, Samples in Ch. 2–12). The system auto-detects active cells.
STEP 3
Auto-Subtract Read
Tap "Auto-subtract" (or "Sample") to initiate reading. The system deducts the blank control automatically and computes inhibition rates after reaction countdown.
STEP 4
Print & Archive
Review pass/fail compliance judgments on screen. Tap "Print" for instant thermal receipt output or enter "Detection Record" to query and export historical logs.
Reading Modes & Compliance Evaluation Guide:
  • Auto-Subtract Mode: Automatically designates the first inserted channel as the Control (C) blank and remaining active channels as Samples (S), streamlining batch enzyme inhibition testing.

  • Individual Channel Control: Operators can also click individual [C] or [S] icons beside any of the 12 channels to run independent control baselines or single-channel readings.

  • Automated Standard Judgment: Built-in regulatory limits (e.g., inhibition rate < 50% threshold per GB 5009.199) automatically flag samples as compliant (OK) or exceeding limits.

6. Performance & Crucial Operational Controls

Adhere to standard instrument maintenance guidelines to guarantee long-term optical precision:

  • Cuvette Removal Before Moving: Always remove all colorimetric cuvettes from the detection chamber immediately after testing to prevent liquid spills from damaging internal optics during transport.

  • Operating Environment: Install on a level, vibration-free benchtop (away from centrifuges or strong magnetic fields) at 5°C–35°C and relative humidity < 70%, out of direct sunlight.

  • Optical Surface Care: Keep cuvette optical faces clean and free of fingerprints or scratches. Never spray liquid cleaners directly into the detection chamber or ports.

  • Dedicated Power Supply: Use exclusively the supplied 12 V DC / 5 A power adapter connected to a grounded 100–240 V AC outlet.

7. Included Components & Auxiliary Equipment
  • Instrument Package Included: 1× PR12 Rapid Pesticide Residue Detector Main Unit, 1× Power Adapter (12 V DC, 5 A) & Power Cord, Colorimetric Cuvettes, Thermal Printing Paper Roll, User Manual.

  • Compatible Consumables & Accessories (Ordered Separately): FSTest Pesticide Residue Rapid Test Kits (Enzyme Inhibition Reagents), replacement 1 cm colorimetric cuvettes, thermal printer rolls, and USB mouse/keyboard.

Request Demonstration & Purchasing Information
Deploy rapid, high-throughput pesticide residue screening across your agricultural supply chain with the FSTest® PR12 Detector. Product quotations, distributor partnerships, and technical consultations are available upon request.

     Email: service@fstestcorp.com | Website: www.fstestcorp.com
Certification & Validation